for Recall
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over | Device Design |
---|
Roche Diagnostics issued on 12/14/18 'Urgent Medical Device Correction (UMDC) via UPS Ground (receipt required) Consignee Instructions: Manually add the EWCs listed in the Urgent Medical Device Correction (UMDC) on your COBAS INTEGRA 400 plus and cobas c 111 analyzers. If customers run the test combinations requiring needle assignment (See Table on page 3 of UMDC TP-00508), run those in batch mode until the updated TAS has been loaded on the COBAS INTEGRA 400 Plus analyzer. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions. |
---|
656 units |
---|
US Nationwide distribution. |
---|
TPLC Device Report |
---|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes. |
510(K) Database | 510(K)s with Product Code = JJE and Original Applicant = Roche Diagnostics 510(K)s with Product Code = JJE and Original Applicant = ROCHE DIAGNOSTICS CORP.
|
---|